IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Modulation Analyzer, Hewlett-Packard, Model 8901A w/Options 002 and 010.
The manufacturer's manual was used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in text, tables, and
figures.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Frequency modulation:
Specifications
Rates
20 Hz to 10 kHz
150 kHz to 10 MHz
20 Hz to 200 kHz
10 to 1300 MHz
20 Hz to 20 kHz
10 to 1300 MHz
(w/750 s filter)
Deviation
40 kHz pk max
150 kHz to 10 MHz
400 kHz pk max
10 to 1300 MHz
40 kHz pk max
10 to 1300 MHz
(w/750 s filter)