TB 9-6695-263-35
Table 1.
Calibration Description - Continued
Test instrument parameters
Performance specifications
AM calibration:
Specifications
Internal reference
Frequency:
10 MHz
Stability
<1 x 10
-9 per day after 30- day warm-up
Audio filters:
50 Hz high pass (2 pole)
50 Hz nominal
<1% at >200 Hz rates
Flatness
300 Hz high pass (2 pole)
300 Hz nominal
<1%>kHz at rates
Flatness
3 kHz low pass (5 pole)
3 kHz nominal
Flatness
<1% at < kHz rates
15 kHz low pass (5 pole)
15 kHz nominal
Flatness
<1% at <10 kHz rates
20 kHz low pass (9 pole bessel)
>20 kHz
<1% at < kHz rates
Flatness
25 s nominal
high pass)
50 s nominal
75 s nominal
750 s nominal
EQUIPMENT REQUIREMENTS
4. Equipment
Required.
Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this