TB 9-6695-293-50
(5) Front Terminal Offset:
(a) Connect a four-terminal short across the front panel terminals as shown in
(b) After connecting the short, allow 5 minutes for thermal equilibrium.
D
NOTE
Take precautions to prevent thermal changes near four-wire
short. Do not touch short after it is installed. If drafts exist,
cover INPUT terminals/short to minimize thermal changes.
(c) Execute the CAL 0 command by pressing: Blue, AUTO CAL, 0, , , 3, 4, 5,
8, and ENTER keys. This adjustment takes about 5 minutes. When completed, the
multimeter will return to displaying dc voltage measurements.
(d) Set to 10 V RANGE and remove four-terminal short from front terminals.
(6) Dc Gain:
(a) Select DCV FUNCTION and set front panel TERMINALS pushbutton to
FRONT.
(b) Connect dc reference standard 10 V output to INPUT HI, LO, and GUARD
terminals.
10