TB 9-6695-293-50
e.
Repeat d above for multimeter range settings and output
voltages listed in
table 3.
Multimeter will indicate within the limits listed.
Table 3. Dc
Voltage Linearity Test Multimeter Indications
Multimeter
Test instrument
Multimeter indications
range setting
output
(V dc)
(V dc)
(V dc)
Min
Max
10
2
01.999980
02.000020
10
3
02.999972
03.000028
10
4
03.999964
04.000036
10
5
04.999956
05.000044
10
6
05.999948
06.000052
10
7
06.999940
07.000060
10
8
07.999932
08.000068
10
9
08.999924
09.000076
10
10
09.999916
10.000084
100
100
099.99900
100.00100
1000
1000
999.9884
1000.0116
f. Set TI to the minimum output then press RESET pushbutton.
10. Resistance Verification
NOTE
The characterization of the multimeter is required for this
performance check. If ambient temperature drifts more than 1
degree Celsius (from time of characterization) prior to
(Configuration modifications will be required depending on value of input resistance; e.g.,
set 0COMP to ON until 100 kΩ, then set 0COMP to OFF.)