TB 9-6695-293-50
b. Set TI for a 1 mA output.
c. Using output adjustment control, set TI for a 1.0000000 mA (10 digits) indication
on multimeter. TI Error display will be within the limits listed in table 5.
Test instrument
Error display
Limits
Output
( ppm)
1
mA
70
10
mA
70
100
mA
80
160
A
100
d. Set TI for a -1 mA output.
e. Using output adjustment control, set TI for a -1.0000000 mA (10 digits) indication
on multimeter. TI Error display will be within the limits listed in table 5.
f. Repeat technique of
b through e above for values listed in
table 5, section A. TI
Error display will indicate within the listed limits.
h.
Configure multimeter for most accurate dc
voltage measurements.
18