TB 9-6695-293-50
c.
Set TI
for a 1 Ω
output. Multimeter will
indicate within
110 ppm of
displayed
value on TI.
d.
Repeat technique of
c above for settings listed in
table 4. Multimeter will
indicate
within the listed limits.
Test instrument
(ppm of test
output
instrument displayed
value)
110
Ω
1.9
Ω
33
10
31
Ω
19
20
Ω
100
20
Ω
190
15
kΩ
1
15
kΩ
1.9
14
kΩ
10
14
kΩ
19
16
kΩ
100
16
kΩ
190
23
MΩ
1
Ω
Ω
0
50
e. Set TI to STANDBY.
g. Set TI
for a
1.9 MΩ
output. Measure resistance. Resistance measuring system will
indicate within 24 ppm of TI displayed value.
h.
Set TI
for a 10 MΩ
output. Measure resistance. Resistance measuring system will
indicate within +46 ppm of TI displayed value.
PIN: 052435-001