TB 9-6695-293-50
(c) Execute the CAL command specifying the 10 V test report value of dc
reference standard. For example, if the test report value is 10.000001 V press: Blue,
AUTO CAL, 1, 0, ., 0, 0, 0, 0, 0, 1, , , 3, 4, 5, 8 and ENTER keys. This adjustment takes
about 2 minutes and, when completed, multimeter will display dc voltage measurements.
(d) Disconnect dc reference standard from multimeter.
(a) Select the four-wire ohms measurement FUNCTION and enable the offset
compensation by pressing: Blue, OHMF/OHM, OFFSET COMPΩ, , and ENTER keys.
(b) Connect standard resistor No. 1 to multimeter using four-wire technique
and set GUARD pushbutton to the TO LO position
(c) Execute the CAL command specifying the test report value of the standard
resistor. For example, if the test report value is 10.00011 kΩ press: Blue, AUTO CAL, 1,
0, ., 0, 0, 0, 1, 1, E, 3, , , 3, 4, 5, 8, and ENTER keys. This adjustment takes about 12
(d) Disconnect standard resistor and execute ACAL AC by pressing: AUTO
CAL,
, 3, 4, 5, 8, and ENTER keys.
,
WARNING
HIGH VOLTAGE is used or exposed during the performance of
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Connect TI to a 115 V ac source.
NOTE
If TI is being calibrated in conjunction with a 5725A/( )
amplifer, the 5725A/( ) should be connected to TI at this time.
The 5725A/( ) should be warmed up for 30 minutes and in the
Afterwards, the 5725A/( ) could be in the OFF state until
b. Set POWER switch to ON and allow at least 30 minutes to warmup