TB 9-6695-293-50
i.
Set TI
for a 19 MΩ
output. Measure resistance. Resistance measuring system will
indicate within 55 ppm of TI displayed value.
j.
Set TI
for a
100 MΩ
output. Measure resistance. Resistance measuring system will
indicate within 130 ppm of TI displayed value.
k. Press RESET.
NOTE
The characterization of the multimeter is required for this
performance check. If ambient temperature drifts more than 1
degree Celsius (from time of characterization) prior to
a. Connect OUTPUT HI, LO, and GUARD terminals to multimeter INPUT I, LO,
and GUARD terminals. Configure multimeter for most accurate dc current measurements.