TB 9-6695-293-50
Table 8. Ac Voltage, 2 mV through 1000 V - Continued
Test instrument
Measured
Output level
error
Limits
(V)
(%)
(0 %)
10
V
10
Hz
0.070
10
V
30
Hz
0.021
10
V
1
kHz
0.0092
10
V
30
kHz
0.016
10
V
90
kHz
0.032
10
V
200
kHz
0.077
10
V
400
kHz
0.190
10
V
1
MHz
0.390
100
V
10
Hz
0.070
100
V
30
Hz
0.021
100
V
1
kHz
0.010
100
V
30
kHz
0.029
100
V
90
kHz
0.070
100
V
200
kHz
0.270
1000
V
1
kHz
0.0094
14. Wideband Ac
Voltage (Option 03)
(1) Verify that TI has warmed up for at least 30 minutes.
(2) Press SETUP MENUS soft key.
(3) Press CAL soft key; then press CAL again.
(4) Press WIDEBAND GAIN soft key.
(5) Enter the ambient temperature and press ENTER pushbutton.
(6) Connect TI as shown in figure 11 for positive gain calibration.
(7) Press ENTER pushbutton.