TB 9-6695-293-50
TEST INSTRUMENT
OUTPUT
SENSE
VΩA
VΩ
WIDEBAND
HI
HI
LO
LO
HI
V-GUARD GROUND
CABLE AND TERMINATION
SUPPLIED WITH
TEST INSTRUMENT.
NOTE
As wideband gain calibration proceeds, messages appear on the
display identifying all processes as they are encountered.
Proceed to (8) below when positive gains calibration is
complete.
(8) For negative gain calibration, reverse connection at the SENSE input so that
center conductor is connected to LO and then press ENTER pushbutton.
(9) At completion of (8) above, set rear panel CALIBRATION NORMAL/ENABLE
switch to ENABLE and press STORE VALUES soft key.
(10) Press DONE WITH CAL soft key until TI display returns to normal. Then set
rear panel CALIBRATION NORMAL/ENABLE switch to NORMAL.
(11) Press RESET pushbutton.
26