TB 9-6695-293-50
NOTE
In both cases, the TI will have to be connected to the 5700A
and in the ON state.
19. High Ac
Voltage Verification
Press INPUT 2 pushbutton on ac
measurement standard.
AC MEASUREMENT STANDARD
SIDE VIEW
HI
INPUT 2
LO
COAX CABLE
(APPROX. 12 INCHES)
BNC TO
BANANA ADAPTER
CALIBRATOR
SIDE VIEW
HI
OUTPUT
LO
TEST INSTRUMENT
HI CURRENT SIDE VIEW
LO OUTPUT
b. On the ac measurement standard, press the UTIL MENUS pushbutton and then
the MEAS CONTROL soft key. Set DIGITAL FILTER MODE to FAST and RESTART
to MEDIUM. Press the DONE soft key twice to return to the measurement display.
c. Set ac measurement standard to manual 700 V range.
d. Set the TI for a 300 V, 20 kHz output. Adjust TI output adjustment knob for an
indication of 300.000 V (1 digit) on the ac measurement standard.
e.
Record calibrator Error display indication in
table 14. Verify that the
result is
within the limits listed.