TB 9-6695-293-50
b.
Configure multimeter for most accurate dc
voltage measurements.
c. Set TI for a +10.0 A dc output.
d. Allow approximately 10 minutes for stabilization; then record multimeter indication.
e.
Calculate measured value by
dividing multimeter indication by
test report value for
standard resistor No. 4. Resulting value will be between 9.99592 and 10.00408 A (4.99772
and 5.00228); if not, perform b below.
f.
Calculate percent of
error for only
+10 A dc and
record for use in
paragraph 21c
below.
NOTE
The following formula may be used to calculate percent of
error:
Measured Value x 100
-
100 = Error (%)
Nominal Value
g. Press +/- then ENTER pushbutton.
h. Repeat technique of d and e above with a -10.0 A dc output. Resulting value will be
between -999592 and -10.00408 A (-4.99772 and -5.00228).
i. Repeat technique of c through h above for 5 A dc, using values in parenthesis.
j.
Set to STANDBY; then press RESET.
Press INPUT 1 pushbutton on ac
measurement standard.