TB 9-6695-293-50
a. Performance Check
(1) Connect equipment as shown in figure 17. Press INPUT 2 pushbutton on ac
measurement standard.
NOTE
Ensure that OUTPUT LOW and (chassis ground) terminal
connecting strap is securely in place on TI.
(2) Set both TI and calibrator POWER switches to ON.
Allow at least 1 hour for
equipment to warmup and stabilize.
NOTE
The operation of the TI is controlled by the calibrator. The TI
controls referred to in the remainder of this check are located
on the calibrator.
(3) On the ac measurement standard, press the UTIL MENUS pushbutton and
then the Meas Control soft key. Set DIGITAL FILTER MODE to FAST and RESTART
to MEDIUM. Press the DONE soft key twice to return to the measurement display.