TB 9-6695-293-50
Figure 18. Power amplifier -
adjustment locations.
Figure 18. Power amplifier adjustment locations.
a. Performance Check
NOTE
Do not perform power supply check if other parameter is within
tolerance.
(1) Connect digital multimeter between TP11 +15V and TP19 GND (fig. 18). If
digital multimeter does not indicate between 14.95 and 15.05 V, perform b below.
(2) Connect digital multimeter between TP12 -15V and TP19 GND (fig. 18). Digital
multimeter will indicate between -14.65 and -15.35 V.
b.
Adjustments.
Adjust R82 +15V (fig. 18) for a
digital multimeter indication between
+14.95 and 15.05 V dc. (R)
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.