TB 9-6695-293-50
d. Set TI and calibrator POWER switches to ON and allow at least 30 minutes before
proceeding.
e.
Configure multimeter for most accurate dc
voltage measurements.
f. Ensure TI is in STBY mode; multimeter will indicate 0 1 V.
g. Connect a short jumper wire between the front VOLTAGE INPUT HI and LO
terminals.
h.
Set to
OPR mode. Multimeter will
indicate 0
100 V
dc; if
not, adjust R6 on the A5
i. Set to STBY mode and remove jumper VOLTAGE INPUT terminals.
j.
Set TI and calibrator POWER switches to OFF.
k. Connect TI to calibrator, John Fluke, Model 5700A/CT. The calibrator should be the
one normally used with the TI and should have been recently calibrated.
l. Set TI and calibrator POWER switches to ON and allow sufficient time to
restabilize.
46