TB 9-6625-102-24
(7) Configure the TI as listed in (a) through (c) below:
(a) Set rotary switch to
mV.
(b) Press SHIFT key for a
indication.
(c)
Press the RANGE key to the 50 mV range.
(8) Set the calibrator voltage and frequency output to the first values in table 5.
Calibrator
TI
Voltage
Range
Min
Max
(mVRMS)
(kHz)
(mVRMS)
(mVRMS)
(mVRMS)
50
1
50
49.760
50.240
50
10
50
49.610
50.390
50
20
50
49.585
50.415
50
100
50
48.130
51.870
500
1
500
497.75
502.25
500
10
500
497.75
502.25
500
20
500
495.85
504.15
500
100
500
481.30
518.70
1000
1
1000
993.5
1006.5
1000
10
1000
993.5
1006.5
1000
20
1000
988.5
1011.5
1000
100
1000
953.0
1047.0
(9) TI will indicate within the limits of table 5; if not, perform b below.
(10) Repeat technique of (7)(c) through (9) above for remaining TI ranges and
calibrator output voltage and frequency values in table 5.
(11) Set the TI as listed in (a) through (c) below:
(a) Set rotary switch to
V.
(b) Press SHIFT key for a
indication.
(c)
Press the RANGE key to the 5 V range.
(12) Set the calibrator voltage and frequency output to the first values in table 6.
Table 6. AC Voltage Secondary Function.
TI
Calibrator
Range (VRMS)
Voltage (VRMS)
Min (VRMS)
Max (VRMS)
5
1
5
4.9775
5.0225
5
10
5
4.9775
5.0225
5
20
5
4.9585
5.0415
5
100
5
4.8130
5.1870
50
1
50
49.780
50.230
50
10
50
49.780
50.230
50
20
50
49.590
50.420
50
100
50
48.130
51.870
500
1
500
497.75
502.25
500
10
500
497.75
502.25
1000
1
1000
992.0
1008.0
(13) TI will indicate within the limits of table 6; if not, perform b below.