TB 9-6625-2013-50
11. X100Ω
Decade Dial Accuracy
a. Performance Check
(1) Set TI decade dials to 000100.0100. If multimeter indication is not between
100.0065Ω and 100.0135Ω, perform b below.
(2) Repeat technique of (1) above for TI decade dial settings and multimeter
indications listed in table 5. If multimeter indications are not within limits specified at
each setting, perform corresponding adjustment listed in table 5.
Test instrument
Multimeter indications
Adjustments for X100
decade dial
(kΩ)
settings
Min
Max
(R)
000200.0100
.2000040
.2000160
R2
000300.0100
.3000015
.3000185
R3
000400.0100
.3999990
.4000210
R4
000500.0100
.4999965
.5000235
R5
000600.0100
.5999940
.6000260
R6
000700.0100
.6999915
.7000285
R7
000800.0100
.7999890
.8000310
R8
000900.0100
.8999865
.9000335
R9
000X00.0100
.9999840
1.0000360
R10
b. Adjustments. Adjust X100
R1 (fig. 1) for a
100.010Ω
multimeter indication (R).
12. X1 kΩ
Decade Dial Accuracy
a. Performance Check
(1) Set TI decade dials to 001000.0100. If multimeter indication is not between
.9999840 and 1.0000360 kΩ, perform b below.
(2) Repeat technique of (1) above for TI decade dial settings and multimeter
indications listed in table 6. If multimeter indications are not within limits specified at
each setting, perform corresponding adjustment listed in table 6.
Test instrument
Multimeter indications
Adjustments for X1K
decade dial
(kΩ)
settings
Min
Max
(R)
002000.0100
1.9999590
2.0000610
R2
003000.0100
2.9999340
3.0000860
R3
004000.0100
3.9999090
4.0001110
R4
005000.0100
4.9998840
5.0001360
R5
006000.0100
5.9998590
6.0001610
R6
007000.0100
6.9998340
7.0001860
R7
008000.0100
7.9998090
8.0002110
R8
009000.0100
8.9997840
9.0002360
R9
00X000.0100
9.9997590
10.0002610
R10