TB 9-6625-2124-35
(6) Press CW key and vary FREQUENCY VERNIER control to tune TI output
frequency from 50 MHz to 2.4 GHz while observing power meter indication. Power meter
will indicate within 0.9 dBm of value recorded in (5) above; if not, perform b (1) through
(20) below.
(7) Enter data 2.5 GHz on keyboard. Power meter will indicate 0.0 1.3 (1.5) dBm.
Record power meter indication.
Test instrument
Power meter
power display indications
indications (dBm)
(dBm)
Min
Max
8.0
6.5 (6.3)
9.5 (9.7)
7.0
5.5 (5.3)
8.5 (8.7)
6.0
4.5 (4.3)
7.5 (7.7)
5.0
3.5 (3.3)
6.5 (6.7)
4.0
2.5 (2.3)
5.5 (5.7)
3.0
1.5 (1.3)
4.5 (4.7)
2.0
0.5 (0.3)
3.5 (3.7)
1.0
-0.5 (-0.7)
2.5 (2.7)
-1.0
-2.5 (-2.7)
0.5 (0.7)
-2.0
-3.5 (-3.7)
-0.5 (-0.3)
-3.0
-4.5 (-4.7)
-1.5 (-1.3)
-4.0
-5.5 (-5.7)
-2.5 (-2.3)
-5.0
-6.5 (-6.7)
-3.5 (-3.3)
(8) Vary FREQUENCY VERNIER control to tune TI output frequency from 2.5 to
7.0 GHz while observing power meter indication. Power meter will indicate within 0.7
dBm of value recorded in (7) above; if not, perform b (21) through (27) below.
(9) Enter data 7.1 GHz on keyboard. Power meter will indicate 0.0 1.3 (1.5) dBm.
Record power meter indication.
(10) Vary FREQUENCY VERNIER control to tune TI output frequency from 7.1 to
13.5 GHz. Power meter will indicate within 0.7 dBm of value recorded in (9) above; if not,
perform b (21) through (27) below.
(11) Enter data 13.6 GHz on keyboard. Power meter will indicate 0.0 1.4 (1.6) dBm.
Record power meter indication.
(12) Vary FREQUENCY VERNIER control to tune TI output frequency from 13.6 to
18.0 GHz. Power meter will indicate within 0.8 dBm of value recorded in (11) above; if
not, perform b (21) through (27) below.
(13) Disconnect power meter from RF OUTPUT.
(14) Press function keys and enter corresponding data on keyboard as listed in (a)
through (e) below:
(a)
INSTR PRESET key.
(b)
START key, DATA ENTRY 50 MHz.
(c)
STOP key, DATA ENTRY 18 GHz.
(d)
POWER LEVEL key, DATA ENTRY -1 dBm.
(e)
SWEEP TIME key, DATA ENTRY 100 s.
33