TB 9-6625-2296-35
(30) Press DATA ENTRY ENTER and DATA ENTRY keys to enter value displayed
on measuring receiver in 100 kHz(+) data field. Press DATA ENTRY ENTER key.
(31) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in 100 kHz(-) data field, and press DATA ENTRY
ENTER key.
(32) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
23. Amplitude Modulation Meter
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
(2) Set signal generator for a 121.1 MHz, 0 dB output and turn RF OUTPUT off.
Connect signal generator RF OUTPUT to power splitter input.
(3) Connect one power splitter output to TI ANTENNA and the other splitter
output to measuring receiver sensor module.
(4) Set measuring receiver to measure AM with PEAK/2 detector selected.
NOTE
Pressing More (F6) will scroll various options on the bottom of
the display. The options will change the functions assigned to
the F1 through F6 function keys. This may be necessary to
obtain the Freq (F1) and Atten (F4) options required in the
following tests.
(5) Press MODE RCVR (Receiver Display), Freq (F1) (Frequency), DATA ENTRY
1 2 1 . 1 and M keys (121.1 MHz).
(6) Press Atten (F4) (Attenuation), DATA SCROLL arrow keys to select 40 dB,
and press DATA ENTRY ENTER (40 dB) key.
(7) Press SETUP (MODE RCVR Menu), DATA ENTRY 2 (Select Mod), DATA
SCROLL arrow keys to select 11. User Defined, and DATA ENTRY ENTER key (to
select User Defined Mod).