Table 15. Frequency Readout Accuracy 9 GHz
Table 17. Frequency Span Accuracy

TB-9-6625-2333-35 Spectrum Analyzer Agilent Models 8562EC-104 and 8562EC-104 Manual
Page Navigation
  23    24    25    26    27  28  29    30    31    32    33  
TB 9-6625-2333-35
Figure 4. Frequency span accuracy - equipment setup.
(2) Press TI INSTRUMENT STATE PRESET key.
(3) Set signal generator frequency to 300.0008 MHz and level output to -15 dBm.
(4) Press TI keys and enter values using DATA keys as listed in (a) through
(g) below:
(a)
FREQUENCY then [CENTER FREQ] to 300.0004 MHz.
(b)
SPAN then [SPAN] to 1 kHz.
(c)
AMPLITUDE then [REF LEVEL] to -10 dBm.
(d)
INSTRUMENT STATE SGL SWP wait for completion of new sweep.
(e)
MARKER PEAK SEARCH.
(f)
[MARKER DELTA].
(g)
[NEXT PEAK].
NOTE
The active and anchor markers should be on the signals near
the second and tenth graticule lines (the left most graticule is
the first graticule line on the display).
(5) TI ∆ MKR indication will be between 792 and 808 Hz.
(6) Press TI MARKER MKR key then [MARKERS OFF] key.
(7) Repeat technique of (3) through (6) above for signal generator and TI settings
listed in table 17. TI ∆ MKR indications will be within limits specified in table 17.
28