Figure 4. Frequency span accuracy - equipment setup.
Figure 5. Frequency response - equipment setup. - TB-9-6625-2333-350030

TB-9-6625-2333-35 Spectrum Analyzer Agilent Models 8562EC-104 and 8562EC-104 Manual
Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
TB 9-6625-2333-35
Table 17. Frequency Span Accuracy
Signal generator
Test instrument
CENTER FREQ
∆ MKR
Frequency
SPAN
settings
settings
indications
settings
(MHz)
(MHz)
Min
Max
300.0016
300.0008
2 kHz
1.584 kHz
1.616 kHz
300.004
300.002
5 kHz
3.96  kHz
4.04  kHz
300.008
300.004
10 kHz
7.92  kHz
8.08  kHz
300.016
300.008
20 kHz
15.84  kHz
16.16  kHz
300.04
300.02
50 kHz
39.6
kHz
40.4
kHz
300.08
300.04
100 kHz
79.2
kHz
80.8
kHz
300.16
300.08
200 kHz
158.4
kHz
161.6
kHz
300.4
300.2
500 kHz
396
kHz
404
kHz
300.8
300.4
1 MHz
792
kHz
808
kHz
301.6
300.8
2 MHz
1.584 MHz
1.616 MHz
304
302
5 MHz
3.8
MHz
4.2
MHz
308
304
10 MHz
7.6
MHz
8.4
MHz
316
308
20 MHz
15.2
MHz
16.8
MHz
340
320
50 MHz
38
MHz
42
MHz
380
340
100 MHz
76
MHz
84
MHz
460
380
200 MHz
152
MHz
168
MHz
700
500
500 MHz
380
MHz
420
MHz
1100
700
1 GHz
760
MHz
840
MHz
1900
1100
2 GHz
1.52  GHz
1.68  GHz
b. Adjustments. Refer to paragraph 6. c. above.
16. Frequency Response
a. Performance Check
NOTE
Perform measuring receiver and sensor module (11792A) ZERO
and CALIBRATE.
(1) Connect equipment as shown in figure 5.