TB 9-6625-2341-24
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE which will display/highlight CONTR.
(4) RESET calibrator.
(5) Set measure to V DC by pressing (a) through (d) below:
(a) MEASURE MENU to display/highlight (V, Hz, dB, TIME).
(b) (F2) V.
(c) Toggle
or
to display/highlight V DC METER.
(d) (F5) SELECT ITEM.
(6) Select display change by pressing (a) through (h) below:
(a)
SUB MENU.
(b)
(F1) MORE METER.
(c)
Toggle
or
to display/highlight READINGS on DISPLAY.
(d) (F5) SELECT ITEM which will display/highlight (1, 2, 3, 4).
(e)
Toggle
or
to display/highlight 1.
(f)
(F5) SELECT ITEM.
(g)
(F1) CLOSE.
(h) SCOPEMETER/MAIN MENU to display SCOPE/METER/:/EXT.mV and etc.
(7) Perform steps as listed in (a) through (d) below for each row in table 5.
(a) Under INPUT A pushbutton left side, toggle (mV RANGE V) to display TI
Range settings.
(b)
Set calibrator to Output voltage settings.
(c)
If TI Voltage indications are not within specified limits, perform b below.
(d)
Set calibrator to standby.
Calibrator
Test instrument
Voltage indication
(DC)
Output volt settings
Range settings
Min
Max
(DC)
(DC)
.1 V
100
mV
99.00 mV
101.00 mV
.3 V
300
mV
298.0
mV
302.0
mV
1
V
1
V
0.990 V
1.010 V
3
V
3
V
2.980 V
3.020 V
10
V
10
V
9.900 V
10.10 V
30
V
30
V
29.80 V
30.20 V
100
V
100
V
099.0
V
101.0
V
300
V
300
V
298.0
V
302.0
V
b. Adjustments. Complete adjustments are in paragraph 16 (R).
20