DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
Headquarters, Department of the Army, Washington, DC
12 December 1985
TB 11-6625-2713-50, 29 December 1978, is changed as follows:
Page 1, Reporting of Errors. Change address to read "Commander, U. S. Army TMDE
Support Group, ATTN: AMXTM-LPE, Redstone Arsenal, AL 35898-5400."
Paragraph 2a. In
lines 3 and 5,
change "TM
38-750" to
read "TB
750-25."
(7) Set TI VERTICAL MODE switch to CH2 and connect digital voltmeter (A3) to
junction of A2A3R9 and A2A3R20 (fig. 2) and chassis ground. If digital voltmeter does not
indicate zero voltage 50 mV, perform b(1) below.
Page 6, table 4. In line 5 u nder maximum
column, change "0.22846" to
read "0.28846."
b. When all parameters are within tolerance, annotate and affix DA Label 80 (U.S.
Army Calibrated Instrument). When the TI receives limited or special calibration,
annotate and affix DA Label 163 (U.S. Army Limited or Special Calibration). When the TI
cannot be adjusted within tolerance, repair the TI in accordance with the maintenance
manual. When repair is delayed for any reason or the TI cannot be repaired with local
resources, annotate and affix DA Form 2417 (U.S. Army Calibration System Rejected
Instrument) and inform the owner/user accordingly in accordance with TB 750-25.
PIN: 015255-002