Figure 9. MODE 4 SLS pulse setup.
APPENDIX A. RADAR TEST SET MENUS

Calibration Procedure For Transponder Set, Test Set An/Apm-378 Page Navigation
  12    13    14    15    16  17  18  
img
. . . . . .
TB 9-4920-358-24
18. Power Supply
a. Performance Check
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
(1) Connect multimeter - (negative) lead to TI TP1 (fig. 2) and + (positive) lead to
chassis. Multimeter indication will be between -50 and +50 mV.
(2) Move multimeter + (positive) lead to TI TP3 (fig. 2). Multimeter indication will
be between 4.75 and 5.25 Vdc.
(3) Move multimeter + (positive) lead to TI TP2 (fig. 2). Multimeter indication will
be between 11.5 and 12.5 Vdc.
b. Adjustments. No adjustments can be made.
19. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective covers on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.