Figure 28. Vt_DC Connection.
Adjustments

Calibration Procedure For Engine Torquemeter Test Set (Etts) Pn 09-1009-02 Page Navigation
  19    20    21    22    23  24  25    26  
img
. . . . .
TB 9-4920-467-24
(3)  The preferred voltage output of +17.336 V dc for Vt_DC is shown in the output
field on the UxValidator. Using the + and - buttons on the screen, adjust the TI's output so
that the voltage displayed on the multimeter(s) matches the on-screen voltage.  The
resolution of the adjustment can be changed by selecting Fine, Medium, or Coarse.
(4)  When the voltage displayed on the multimeter matches the voltage displayed in
the output field on the UxValidator, Vt_DC is properly adjusted.
(5)  Select and activate Done (in the upper right corner of the screen) on the
UxValidator.
Repeat paragraph 19 a above.
(6)
20. Vc_DC/Vc_10
a.
Performance Check
(1)  Select and activate Vc_DC from the Circuit menu on the UxValidator. Vc_DC
will appear in the Circuit field.
(2)
Connect equipment as shown in figure 30.
Set both resistance standards to 32.4 kΩ.
(3)
Figure 30. Vc_DC Connection.
(4)
Select and activate Test from the POD Alignment screen on the UxValidator.