(5) Connect equipment as shown in figure 28, connection B. If reading on
multimeter does not indicate between 17.321 and 17.351 V dc, perform b below.
(6) Press the TI ETTS BrainPak left Encoder button to return to the main POD
Alignment screen.
b.
Adjustments
(1)
Connect equipment as shown in figure 28, connection B.
(2) Select and activate Align on the UxValidator. Several fields on the display will
change color as shown in figure 29. The multimeter will display the Vt_DC stimulus output
(between +16 and +18 V dc).
24