Figure 23. VoCmp Connection.
Figure 26. Jrt Connection.

Calibration Procedure For Engine Torquemeter Test Set (Etts) Pn 09-1009-02 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . . . . .
TB 9-4920-467-24
b.
Adjustments
(1)  If the performance check fails, perform the adjustment procedure for Vc in
paragraph 11 b first, then the performance check for Vc in paragraph 11 a.
(2)  Repeat until the performance check passes, then continue with paragraphs 12
through 15.
16. Jrtc
a.
Performance Check
(1)  Select and activate Jrtc from the Circuit menu on the UxValidator. Jrtc will
appear in the Circuit field.
(2)
Connect resistance standard to TI ETTS Pod as shown in figure 25.
(3)
Select and activate Test from the POD Alignment screen on the UxValidator.
(4)  Set the resistance standard to 900.00 Ω. If TI reading on UxValidator does not
indicate between 899.00 and 901.00 Ω, perform b below.
(5)  Set the resistance standard to 1000.00 Ω.
If TI reading does not indicate
between 999.00 and 1001.00 Ω, perform b below.
(6)  Set the resistance standard to 1100.00 Ω.
If TI reading does not indicate
between 1099.00 and 1101.00 Ω, perform b below.
(7)  Press the TI ETTS BrainPak left Encoder button to return to the main POD
Alignment screen.
Figure 25. Jrtc Connection.
b.
Adjustments
(1)  Select and activate Align on the UxValidator.
On-screen instructions are
displayed.