Figure 2. Equipment Setup (Model TT1200A).

Calibration Procedure For Turbine Temperature Test Set Models Tt1000A, Tt1200 And Tt1200A Page Navigation
  11    12    13    14    15  16
img
. . .
TB 9-4920-469-24
(71)Short the clips together. (Ensure they are properly shorted together).
(72)The SYS RES will now be tested.
(73)Set the RANGE switch to 200Ω.
(74)Press PTS and hold until the TI display settles.
(75)The TI displays a value less than 2.00 Ω.
(76)Turn the SYS RES knob fully clockwise (CW).
(77)Press and hold in the PTS button until the TI display stabilizes.
(78)The TI displays a value greater than 25.00 Ω.
NOTE
If the TI fails the test, an error message displays.  The TI
displays a message (DBLR CAL ENT/ESC) to continue to the
mV doubler calibration test or to skip test.
(79)Press CONT (ESC) to skip the test.
(80)Repeat paragraphs 9 and 11 above.
13. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.