Table 1. Calibration Description - Continued.
SECTION III. CALIBRATION PROCESS

Calibration Procedure For Signal Generator, Sg1112Viu, Sg1112V2U, And Mis-28707Ty2 Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . . . . .
TB 9-4931-488-24
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-440. Alternate items may be used
by the calibrating activity when the equipment listed in table 2 is not available. The items
selected must be verified to perform satisfactorily prior to use and must bear evidence of
current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parentheses.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.  The following peculiar accessories are also required for this
calibration: Crystal Detector, HP 432A and a 600 Ω termination.
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
Distortion:
AUDIO ANALYZER
Boonton, Model 1121 (1121)
Range:0 to 20% distortion
Accuracy: 3% FS
Frequency:
Range:90 to 100 kHz
Accuracy: <1% distortion
AUTOTRANSFORMER
Range: 105 to 123 V dc
Ridge, Model 9020A (9020A)
Accuracy:  1%
FREQUENCY COUNTER
Range: 20 Hz to 1100 MHz
Fluke, Model PM6681/656
-8
(PM6681/656)
Accuracy:   1.25 x 10
MEASURING RECEIVER
Range: -112 to + 20 dBm
Measuring receiver system N550S
consisting of Spectrum Analyzer,
Accuracy:  0.37 dB
Agilent Model E4440A (E4440A),
Range: 9.5 to 96.5% (AM)
Power meter, Agilent Model
Accuracy:  1% (AM)
E4419B (E4419B) or N1911A
Range:  4.7 to 339 kHz (FM)
(N1911A), and Sensor module,
Accuracy:  1.5% (FM)
Agilent Model 504 (504)
Hewlett Packard, Model 3458A,
MULTIMETER
Range: 0.196 to 4.08 V ac
(3458A)
Accuracy:  0.5%
Range: 0.000 to 44.70 V dc
Accuracy:  0.025%
OSCILLOSCOPE
Bandwidth: 50 MHz
Agilent, OS-303/G
Sensitivity: 5 mV/div
(OS-303/G)
Accuracy:  3%
PULSE GENERATOR
Range: 1 kHz rate, 0.5 ms width
Agilent, Model 81150A Pulse
Function Arbitrary Noise
Generator (81150A)