Table 18. Internal Modulation Oscillator Frequency Accuracy and Distortion.
Sensitivity, Accuracy, and Distortion (All Models Except Those Containing

Calibration Procedure For Signal Generator, Sg1112Viu, Sg1112V2U, And Mis-28707Ty2 Page Navigation
  20    21    22    23    24  25  26    27    28    29    30  
img
. . . . . . . . . . .
TB 9-4931-488-24
(3) Set MODULATION FREQUENCY range switch to X100 and adjust
MODULATION FREQUENCY vernier dial fully ccw.
(4) Connect true rms voltmeter INPUT to A11TP4 (fig. 3) and chassis ground.
Adjust A11R28 (fig. 3) for a 1.4 (1.37 to 1.48) V ac true rms voltmeter indication. Record
true rms voltmeter indication.
(5) Connect frequency counter CHANNEL A input to TI AM OUTPUT. Adjust
MODULATION FREQUENCY vernier dial fully cw.
(6) Adjust A11C2 and A11C3 (fig. 3) until true rms voltmeter indicates within
0.1 V ac of value recorded in (4) above.
NOTE
Adjusting A11C2 (fig. 3) ccw decreases voltage while increasing
frequency. Adjusting A11C3 (fig. 3) ccw increases both voltage
and frequency.
(7) Adjust MODULATION FREQUENCY vernier dial to 20. If frequency counter
does not indicate 2.00 0.02 kHz, adjust vernier dial for a 2.00 kHz frequency counter
indication. Loosen setscrews in vernier control shaft gear and position vernier dial to 20.
Tighten setscrews and, if necessary, readjust A11R28 (fig. 3) for true rms voltmeter
indication between 1.37 and 1.48 V ac. Record true rms voltmeter indication.
(8) Adjust MODULATION FREQUENCY vernier dial to 200.  If true rms
voltmeter does not indicate within 0.01 V ac of value recorded in (7) and frequency counter
indication is not 20 0.2 kHz, readjust A11C2 (fig. 3) and A11C3 (fig. 3) for desired
indications.
(9) Adjust MODULATION FREQUENCY vernier dial to 20. If true rms voltmeter
does not indicate within 0.01 V ac of value recorded in (7) above and frequency counter
indication is not 2.00 0.02 kHz, repeat technique of (7) and (8) above.
(10) Adjust MODULATION FREQUENCY vernier dial to 20. True rms voltmeter
should indicate between 1.37 and 1.48 V ac. Record true rms voltmeter indication.
(11) While observing true rms voltmeter indication, slowly vary TI modulation
frequency from 2 kHz to 20 kHz. True rms voltmeter should indicate within 0.03 V ac of
indication recorded in (10) above.
(12) Set  MODULATION  FREQUENCY  range  switch  to  x3k  and  adjust
MODULATION FREQUENCY vernier dial to 200. Adjust A11C9 (fig. 3) for a 600 6 kHz
frequency counter indication.
(13) Position controls as listed in (a) through (c) below.
(a) FM switch to INT.
(b) MODULATION FREQUENCY range switch to x100.
(c) MODULATION FREQUENCY vernier dial to 20.
(14) Connect true rms voltmeter INPUT to A11TP5 (AM OUT, fig. 3) and chassis
ground. Adjust A11R35 (fig. 3) for a 0.840 0.010 V ac true rms voltmeter indication.
Record true rms voltmeter indication.
25