Sensitivity, Accuracy, and Distortion (All Models Except Those Containing
Table 21. AM Distortion.

Calibration Procedure For Signal Generator, Sg1112Viu, Sg1112V2U, And Mis-28707Ty2 Page Navigation
  22    23    24    25    26  27  28    29    30    31    32  
img
. . . .
TB 9-4931-488-24
Figure 5. AM - Equipment Setup.
(3) Set measuring receiver to acquire the frequency then to measure AM with peak
+ detector. Set high-pass filter to 50 Hz and low-pass filter to 15 kHz.
(4) Set pulse generator controls for a 1 kHz continuous sine wave output frequency
and 600 Ω load impedance.
NOTE
In techniques (5) and (7) below, record measuring receiver
% AM display indication using Peak +/- 2 detector.
(5) Adjust pulse generator output controls for levels in table 19. If the measuring
receiver does not indicate between min and max, perform b below.
Table 19. AM Sensitivity.
Measuring receiver % AM
Pulse generator
indications (%)
Output level (V rms)
Min
Max
0.636
85.5
94.5
0.495
66.5
73.5
0.354
47.5
52.5
0.212
28.5
31.5
0.141
19.0
21.0
0.071
9.5
10.5
(6)
Adjust pulse generator output controls for a 1.414 V rms level.
(7) Adjust AM control for TI meter indications as listed in table 20. The measuring
receiver should indicate between min and max limits.
27