Equipment Setup
Table 4. Bandwidth Check

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . .
TB 9-4931-538-40
(a) Press Display key. Display Mode to Averaged is highlighted. NUMBER
OF AVERAGES to # Averages = 64 is highlighted. Screen to Single is highlighted.
Graticule to GRID is highlighted. Bandwidth to 12.4 GHz is highlighted.
(b) Press Channels key. Channel 1 display On is highlighted. Channel 2
through 4 display Off is highlighted. VOLTS/DIV to Sensitivity 80.00 mV/div.
(c) Press Timebase key and set TIME/DIV to Sweep Speed = 5.00 ns/ div.
(4) Connect equipment as shown in figure 1, CONNECTION A.
Figure 1. Low bandwidth - equipment setup.
(5) Set synthesized signal generator amplitude for 12.0 dBm and frequency of 50 MHz.
(6) Measure and record the mV reading of the synthesized signal generator and
measuring receiver in table 4, part 1.
6