TB 9-5120-212-35
NOTE
On some models, it is necessary to lift pin on indexer ratchet
drive on top of TI to change cw and ccw directions.
Test instrument
Torque cell indications (ft-lbs)
indications (ft-lbs)
Min
Max
Model PD1201 (and
Models DPT-1200 and DPT-1200R)
(240)
(235.2)
(244.8)
300 (400)
276
(392)
324
(408)
500 (600)
476
(588)
524
(612)
750 (800)
726
(784)
774
(816)
1000 (1000)
976
(980)
1024
(1020)
1200 (1200)
1176
(1176)
1224
(1224)
Model PD2501 (and
Models DPT-2500 and DPT-2500R)
1500
1450 (1470)
1550 (1530)
2000
1950 (1960)
2050 (2040)
2500
2450 (2450)
2550 (2550)
Model PD 12003
5500
5335
5665
6000
5820
6180
NOTE
Perform (4) below for ccw calibration only when specifically
requested by user.
(4) Perform paragraph 7 above and select the torque cell for the TI range and cw
(5) Repeat a(1) through (3) above, except turn crank handle ccw.
(6) Turn TI crank handle cw to obtain a 0 indication.
b. Adjustments
NOTE
Reaction arm must be free to move under retaining screws (fig. 1).
6 CHANGE 1