TB 9-5210-215-24
Figure 2. Test Instrument Orientation (0 to 45).
Table 3. Angle Indication 0 to 45.
Test instrument2
Sine plate
Gage block
Gage block
Indication (o)
selection1
angle
Position
stack height
(o)
(inch)
(inch)
Min
Max
0
0
N/A
1
-0.1 (-0.05)
0.1 (0.05)
0
0
2
-0.1 (-0.05)
0.1 (0.05)
0
0
3
-0.1 (-0.05)
0.1 (0.05)
0
0
4
-0.1 (-0.05)
0.1 (0.05)
5
0.4358
0.1008
1
4.9 (4.95)
5.1 (5.05)
0.135
5
0.4358
2
4.9 (4.95)
5.1 (5.05)
0.2
5
0.4358
3
4.9 (4.95)
5.1 (5.05)
5
0.4358
4
4.9 (4.95)
5.1 (5.05)
15
1.2941
0.1001
1
14.8
15.2
0.050
15
1.2941
2
14.8
15.2
0.144
15
1.2941
3
14.8
15.2
1
15
1.2941
4
14.8
15.2
30
2.5000
0.5
1
29.8
30.2
2
30
2.5000
2
29.8
30.2
30
2.5000
3
29.8
30.2
30
2.5000
4
29.8
30.2
45
3.5355
0.1005
1
44.8
45.2
0.135
45
3.5355
2
44.8
45.2
0.3
45
3.5355
3
44.8
45.2
3
45
3.5355
4
44.8
45.2
Gage block selection is based on the set in table 2 and may be different for other sets.
1
Indications in parentheses are for model PRO3600.
2
b. Adjustments. Complete all performance checks in the entire procedure noting AS
FOUND failures. Use paragraph 10 for adjustments.