TB 9-5210-215-24
9. Angle Indication Greater Than 45
a. Performance Check
(1) Setup equipment on level surface plate as shown in figure 3 using the 5-inch
sine plate.
Figure 3. Angle Indication (> 45) Equipment Setup.
(2) Ensure TI HOLD and ALT ZERO functions are deactivated.
(3) Check TI in positions 5 through 8 (fig. 4). TI will indicate within limits specified
Figure 4. Test Instrument Orientation (>45).