TB 9-5210-215-24
Figure 5. Test Instrument Superset Alignment Setup.
f. Position TI in position 3 (fig. 2), wait 10 seconds, then press TI HOLD button until
[3] appears.
g. Position TI in position 4 (fig. 2), wait 10 seconds, then press TI HOLD button until
[4] appears.
NOTE
TI should not be allowed to rest on narrow end since this
surface is not calibrated.
h. Position TI as shown in figure 3, wait 10 seconds, then press TI HOLD button until
[5] appears.
i. Position TI in position 6 (fig. 4), wait 10 seconds, then press TI HOLD button until
[6] appears.
j. Position TI in position 7 (fig. 4), wait 10 seconds, then press TI HOLD button until
[7] appears.
k. Position TI in position 8 (fig. 4), wait 10 seconds, then press TI HOLD button until
[8] appears briefly, followed by regular measurement.
l.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA equipment label/form in accordance with TB 750-25.