TB 9-5855-1890-24
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 0 to 5 C differential
DIFFERENTIAL
Omega Engineering 0.020
Accuracy: 0.11 C
Type K (7917040)1
THERMOCOUPLE SET
Range: 0 to 200 V
Hewlett Packard, Model 3458A
Accuracy: 2 V
(3458A)
Range: 0 to 6 V dc
Fluke, Model 8840A/AF05
Accuracy: 0.2%
(AN/GSM-64D)
1Part
Description
Common name
(part number)
ADAPTER1
Offset voltage pin (7917036)2
COTTON TIP APPLICATOR3
Swap with cotton tip on one end
HOLDING CLAMP1
Mounting post clamp (7917039) 2
LABORATORY CLAMP1
Laboratory clamp (MS-36012-1) 2
42-in., single banana to test hook (red) (7917035-2) 2
LEAD
42-in., single banana to test hook (black) (7917035-1) 2
LEAD
PASTE REMOVAL SOLVENT3
WD40 or equivalent
THERMOCOUPLE MOUNTING POST1
Mounting post with base (7917037) 2
THERMALLY CONDUCTIVE PASTE3
Omegatherm 201 or equivalent
1Two
required.
2Component part of calibration kit 7917034, limited deployed.
3Additional equipment required.
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 9-5855-255-14 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.