Figure 1. Temperature controller locations.
Calibration

Calibration Procedure For Thermal Collimator, Sm-D-805691 P/O An/Tam-3 And An/Tam-3A And Infrared Collimator, Sm-D-969722 Page Navigation
  1    2    3  4  5    6    7  
img
. . . . . . . .
TB 9-5855-1890-24
g. Increase tension on thermocouple TC1 (fig. 3) to produce a moderate bow in
thermocouple wire by sliding laboratory clamp forward (toward temperature controller) in
mounting post clamp; then tighten mounting post clamp.
h. If necessary, physically position thermocouple TC1 on heater plate to the position
shown in figure 3.
i. Daub a small amount of thermally conductive paste on side of brass housing of MRT
target housing (fig. 3) (on temperature controller) and place tip of thermocouple TC2 (fig. 3)
in the thermally conductive paste.
j.
Repeat technique of f through h above for thermocouple TC2 (fig. 3).
k. Place leads from differential thermocouple (fig. 2) on base of TI near exit port.
Figure 2. Equipment setup.
Figure 3. Temperature controller
heater plate.
5