The measuring receiver will be used in the Tuned RF Level
mode. Calibration of the measuring receiver at the frequencies
and levels used must be accomplished before the measurement
can made taken.
(42) Configure the measuring receiver for a tuned level measurement at each
(43) Using the blue arrow keys and the rotary knob, edit the TI Actual output until
the measuring receiver indicates the Target value stated. Press Accept Target
(44) Press Next Target.
(45) Repeat (42) through (44) above for all points requiring the use of the measuring
receiver to make level adjustments. When complete, the TI adjustment point title will
indicate AM Depth.
(46) Press Previous Menu, the Leveling Head and Base adjustment are now
complete.
a. Deenergize and disconnect all equipment.
b. Annotate DA Label 163 "SSB Phase Noise not calibrated" and affix in accordance
with TB 750-25.