TB 9-5985-314-24
Table 1. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
3
Wiltron/Anritsu, Model
68369NV
NO. 1
Power output: +8 dBm, r1dB
Flatness: 10 to 50 MHz: r2 dB
.05 to 18 GHz: r0.8 dB
Wiltron/Anritsu, Model 68347M
NO. 2
Power output: +11 dBm, r1dB
Flatness: 10 to 50 MHz: r2 dB
.05 to 18 GHz: r0.8 dB
Hewlett-Packard, Model X910B
4
TERMINATION
(X91OB)
VSWR: 1.015 (max)
Hewlett-Packard, Model P910A
VSWR: 1.02 (max)
(P91OA)
Maury Microwave, Model U301
(U301)
1
VSWR:
1As
charted on calibration report provided by Primary Lab (for secondary reference only).
2
3Not calibrated above 18GHz.
4Select as required: Two each may be required.
CALIBRATION PROCESS FOR DIRECTIONAL COUPLERS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 through 8 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration. Omit
steps for parameters listed in Appendix A with the statement N/A.