Coupling and Directivity Tests - Continued
Coupling and Directivity Tests - Continued

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. .
TB 9-5985-314-24
Figure 2. Equipment setup (18 to 26.5 GHz).
(14) Connect TI into figure 2 equipment setup as listed in (a) through (d) below:
(a) INPUT PORT to POINT A.
(b) SIDEARM PORT to POINT B.
(c) OUTPUT PORT to 50 : termination.
(d) AUX (TEST) SIDEARM PORT (if any) to 50 : termination.