Coupling and Directivity Tests - Continued
Figure 3. Equipment setup (26.5 to 40 GHz).

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  5    6    7    8    9  10  11    12    13    14    15  
img
. .
TB 9-5985-314-24
(22) Determine signal generator No. 1 frequency (LO) required for the desired
measurement by calculating the following equation: LO = ((3xRF)-IF)/2.
EXAMPLE B:
Where: RF = signal generator frequency No. 2 determined in (20) above.
IF = operating frequency of receiver system (1 GHz)
LO = Signal generator No. 1 frequency
LO = ((3xRF) -IF)/2
LO = ((3x9 GHz) -1 GHz)/2
LO = 26 GHz/2
LO = 13 GHz
Signal generator No. 1 frequency would be adjusted to 13 GHz.
(23) Adjust signal generator No.1 controls to frequency determined in (22) above and
adjust RF LEVEL output to 0 dBm.
(24) Connect equipment as shown in figure 3.
NOTE
Ensure equipment is connected as shown in figure 3 and has
completed the 3 hour warm-up before proceeding to (25) below.
NOTE
Use 15% SEARCH MODE on receiver system.
12