Figure 2. Equipment setup (18 to 26.5 GHz).

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. .
TB 9-5985-314-24
NOTE
Ensure equipment is connected as shown in figure 2 and has
completed the 3 hour warm-up before proceeding to (9) below.
(9) Adjust signal generator No.1 frequency controls to test frequency recorded in 8 c
above and RF OUTPUT controls for +3 dBm.
(10) Determine signal generator No. 2 frequency (LO) required for the desired
measurement by calculating the following equation: LO = (RF-IF)/2 where:
RF = test frequency recorded in 8 c above IF = intermediate frequency at
which receiver system is to perform measurement (0.700 GHz).
EXAMPLE:
Let IF = 0.700 GHz
RF = 18 GHz (test frequency 8 c above)
LO = signal generator frequency
LO = (RF-IF)/2
LO = (18 GHz-0.700 GHz)/2 LO = 17.3 GHz/2
LO = 8.65 GHz
Signal generator No. 2 frequency for this measurement would be 8.65 GHz.
NOTE
Use receiver system in 15% SEARCH MODE.
(11) Press receiver system frequency to 0.700 GHz.
(12) Adjust signal generator frequency No. 2 to value determined in (10) above and
RF OUTPUT to +8 dBm.
(13) Connect POINT A to POINT B (fig. 2) and establish a 0.00 dBm reference on
receiver system.