Equipment Setup
Coupling and Directivity Tests

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . .
TB 9-5985-314-24
Figure 1. Equipment setup (10 MHz to 18 GHz).
9. Coupling and Directivity Tests
a. Performance Check
NOTE
Perform (1) through (8) below for TI test frequencies from 10
MHz to 18 GHz.
Perform (9) through (19) below for TI test frequencies from 18
to 26.5 GHz.
Perform (20) through (32) below for TI test frequencies from
26.5 to 40 GHz.
(1) Adjust signal generator No. 1 frequency controls to test frequency recorded in
8 c above and adjust RF OUTPUT controls for +6 dBm.
NOTE
Use 15 percent search on receiver system.