Coupling and Directivity Tests - Continued
Coupling and Directivity Tests - Continued

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . .
TB 9-5985-314-24
Figure 3. Equipment setup (26.5 to 40 GHz).
(25) Press receiver frequency to 1 GHz.
(26) Connect POINT A to POINT B (fig. 3) and establish a 0.00 dB reference on
receiver system.
(27) Connect TI into figure 3 equipment setup as listed in (a) through (d) below:
(a) INPUT PORT to POINT A.
(b) SIDEARM PORT to POINT B.
13