TB 9-5985-318-24
(28)Set the power meter for measurement on the 35.0 dBm (0.3 W) range, or for
auto ranging, as applicable.
(29)Verify that the power meter indication is between 37.0 and 43.0 dBm (or
between 50 and 200 nW).
(30)Set the synthesized signal generator RF output to off and disconnect the
equipment from the TI.
(31)Repeat steps (16) through (30) above, except set the synthesized signal generator
for a 1000 MHz CW output.
(32)Unless other measurements are to be performed, turn all power switches to off
or standby, and disconnect the equipment from the TI.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.