• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION II. EQUIPMENT REQUIREMENTS

Calibration Procedure For Rlc Digibridge Model 1693 Page Navigation
1  2    3    4    5    6  
img
. . . . . . .
TB 9-6625-100-24
SECTION I
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of RLC Digibridge IET Labs, Model 1693. The manufacturer's manual was used
as the prime data source in compiling these instructions. The equipment being calibrated
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description.
Test instrument parameters
Performance specifications
Range 1: 25.6 kΩ to 410 kΩ
Resistance
Frequency: 12 Hz to 20 kHz
Basic Accuracy2: 0.02%
Range 2: 1.6 kΩ 25.6 kΩ
Frequency: 12 Hz to 200 kHz
Basic Accuracy2: 0.02%
Range 3: 100 Ω 1.6 kΩ
Frequency: 12 Hz to 200 kHz
Basic Accuracy2: 0.02%
Range 4: 6.25 Ω to 100 Ω
Frequency: 12 Hz to 200 kHz
Basic Accuracy2: 0.02%
Range 1: 25.6 kΩ to 410 kΩ
Capacitance1
Frequency: 12 Hz to 20 kHz
Basic Accuracy2: 0.02%
Range 2: 1.6 kΩ 25.6 kΩ
Frequency: 12 Hz to 200 kHz
Basic Accuracy2: 0.02%
Range 3: 100 Ω 1.6kΩ
Frequency: 12 Hz to 200 kHz
Basic Accuracy2: 0.02%
Range 4: 6.25 Ω to 100 Ω
Frequency: 12 Hz to 200 kHz
Basic Accuracy2: 0.02%
See footnotes at end of table.
2


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.