Table 2. Minimum Specifications of Equipment Required.

Calibration Procedure For Decade Capacitor, Model Hacs-Z-B-6E-1Pf And Hacs-Z-B-3E-1Uf Page Navigation
1  2    3    4    5    6  
img
. . . . . . .
TB 9-6625-113-24
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Decade Capacitor, IET, Model HACS-Z-B-6E-1pF and HACS-Z-B-3E-1F. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. Variations among models are described in text, tables, and figures.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description.
Test instrument parameters
Performance specifications
Capacitance
Zero Capacitance ≤ 0.1 pF
Model: HACS-Z-B-6E-1pF
Range: 1 pF to 1.111,10 F
Frequency: 1 kHz
Accuracy: (0.1%+ 1.0 pF)
Zero Capacitance ≤ 0.5 pF
Model: HACS-Z-B-3E-1F
Range: 1 F Step
Accuracy: 0.1%
Frequency: 1 kHz
Range: 10 F Step
Accuracy: 0.1%
Frequency: 100 Hz
Range: 100 F Step
Accuracy: 0.5%
Frequency: 100 Hz
Range: 1000 F Step
Accuracy: 10%
Frequency: 100 Hz
2