SECTION III. CALIBRATION PROCESS FOR DECADE CAPACITOR IET MODEL HACS-Z-B-6E-1PF
Figure 2. Capacitance Measure HACS-Z-B-6E-1pF.

Calibration Procedure For Decade Capacitor, Model Hacs-Z-B-6E-1Pf And Hacs-Z-B-3E-1Uf Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . .
TB 9-6625-113-24
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel  fail  to  observe  safety  precautions.  REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Measurements can be adversely affected by temperature.
The optimum
temperature for this calibration is 23 C (73.4 F).
b. Exercise TI dials before beginning calibration.
c. Record measurement results for test report if required.
See paragraph 9 for
explanation.
8. Capacitance Accuracy
a. Performance Check
(1) Ensure that the impedance measuring system is set to measure series
capacitance at a test voltage of 1 V rms and a test frequency of 1 kHz.
(2) Perform short and open compensation on impedance measuring system (fig. 1,
connections A and B).
Figure 1. Short and Open Compensation (Coaxial).
5