TB 9-6625-113-24
Table 2. Minimum Specifications of Equipment Required Continued.
Manufacturer and
model
Common name
Minimum use specifications
(part number)
Range: 100 pF to 1000 pF
Frequency: 1 kHz
Accuracy: See chart below.
0.3
0.2
0.1
0
100 200 300 400 500 600 700 800 900 1000
Pico Farads
Range: 1000 pF to 10 F
Frequency: 1 kHz
Accuracy: 0.025%
Range: 10 F to 100 F
Accuracy: 0.025% (0.046%)
Range: 100 F to 1,000 F
Accuracy: 0.125%
Range: 1000 F to 10,000 F
Accuracy: 2.5%
IET MODEL HACS-Z-B-6E-1PF
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.
d. Unless otherwise specified, all controls and control settings refer to the TI.